Procedural Instructions on the TSC
8
8 - 6
Infusomat® fmS 2.3 gb
2.
Document value.
3.
Pull off the MFC service connector.
WARNING
SWITCH THE DEVICE OFF AFTER THE SAFETY CLAMP WAS
CHECKED SO THAT THE ELECTRONIC PRESSURE CONTROL IS RE-
ACTIVATED AFTER A RESTART.
4.
Switch device off.
Delivery Accuracy
Requirements:
-
Mount test setup as shown in
-
Ambient temperature and delivery rate according to the TSC.
1.
Insert line of the test setup in the device.
2.
Vent test setup. Bottom part of drop chamber must be 2/3
filled.
3.
Insert cannula in empty graduated cylinder.
4.
Set delivery rate according to the TSC.
5.
Simultaneously start infusion and stop watch.
6.
Stop the stop watch when the measured volume defined in
the TSC is reached.
7.
Evaluate deviation in percent according to
and
record.
Fig.: 8 - 6
25 ml ± 0.4 ml
Reference height
80 ± 2 cm
Measuring Time
Deviation
%
6 min 40.0 sec
-10
6 min 18.9 sec
-5
6 min 15.0 sec
-4
6 min 11.1 sec
-3
6 min 7.3 sec
-2
6 min 3.6 sec
-1
6 min 0.0 sec
0
5 min 56.4 sec
1
5 min 52.9 sec
2
5 min 49.5 sec
3
5 min 46.2 sec
4
5 min 42.9 sec
5
5 min 27.3 sec
10
Table 8 - 1
Содержание 871 5416
Страница 1: ...Service Manual Infusomat fmS 0 Version 2 3 English 0 10 ...
Страница 10: ...Contact Persons 0 0 10 Infusomat fmS 2 3 gb For your notes ...
Страница 22: ...System Overview 1 1 12 Infusomat fmS 2 3 gb For your notes ...
Страница 30: ...Software 2 2 8 Infusomat fmS 2 3 gb For your notes ...
Страница 60: ...Unit Elements 4 4 20 Infusomat fmS 2 3 gb For your notes ...
Страница 64: ...Maintenance 6 6 2 Infusomat fmS 2 3 gb For your notes ...
Страница 78: ...Test Equipment and Special Tools 9 9 2 Infusomat fmS 2 3 gb For your notes ...
Страница 82: ...Spare Parts List 10 10 4 Infusomat fmS 2 3 gb For your notes ...
Страница 84: ...Index 11 11 2 Infusomat fmS 2 3 gb ...
Страница 86: ...Revision Documentation A A 2 Infusomat fmS 2 3 gb For your notes ...