Service Program
3
Infusomat® fmS 2.3 gb
3 - 7
The activation of the "10 minutes off” alarm is only permissible, if:
-
the staff call is connected and
-
the Infusomat® fmS has an attention label (label drawing no.
M007100000F04).
1.
OK activates the function.
2.
Select alarm tone with NEXT.
3.
Acknowledge with YES.
4.
Return to initial function with END.
Staff Call
Function 430.0
Different staff call modes can be selected:
-
Dynamic with OFF Alarm
-
Dynamic without OFF Alarm
-
Static without OFF Alarm
For further details see staff call line in the instructions for use.
An additional switch-on pulse (YES/NO) can be activated for each
mode to test the staff call unit.
1.
OK activates the function.
2.
Select staff call type with NEXT.
3.
Acknowledge with YES.
4.
Return to initial function with END.
Special Functions
Function 440.0
Special functions can be activated in the Service Program, which
are then available on the user interface. Deactivated special func-
tions will not be displayed. The SM softkey will not be displayed
in standard operation, if all special functions are deactivated. -
Special functions to be selected, see
.
1.
OK activates the function.
2.
Select special functions with NEXT.
3.
Activate / deactivate the respective function with YES/NO.
4.
Return to main menu with END.
Fig.: 3 - 6
Содержание 871 5416
Страница 1: ...Service Manual Infusomat fmS 0 Version 2 3 English 0 10 ...
Страница 10: ...Contact Persons 0 0 10 Infusomat fmS 2 3 gb For your notes ...
Страница 22: ...System Overview 1 1 12 Infusomat fmS 2 3 gb For your notes ...
Страница 30: ...Software 2 2 8 Infusomat fmS 2 3 gb For your notes ...
Страница 60: ...Unit Elements 4 4 20 Infusomat fmS 2 3 gb For your notes ...
Страница 64: ...Maintenance 6 6 2 Infusomat fmS 2 3 gb For your notes ...
Страница 78: ...Test Equipment and Special Tools 9 9 2 Infusomat fmS 2 3 gb For your notes ...
Страница 82: ...Spare Parts List 10 10 4 Infusomat fmS 2 3 gb For your notes ...
Страница 84: ...Index 11 11 2 Infusomat fmS 2 3 gb ...
Страница 86: ...Revision Documentation A A 2 Infusomat fmS 2 3 gb For your notes ...