Unit Elements
4
Infusomat® fmS 2.3 gb
4 - 3
4.4
Pump Cover
Designation
Ord. No.
Pump cover with lock . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 5717
Blind plug 7.1 mm (10 pcs.) . . . . . . . . . . . . . . . . . . . . . 3477 3207
Torsion spring in lever / pump cover (5 pcs.) . . . . . . . 3477 3363
Torsion spring for pump cover (5 pcs.) . . . . . . . . . . . . 3477 3355
Lever (pump cover). . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3477 4092
Hinge pin for pump cover . . . . . . . . . . . . . . . . . . . . . . . 3477 3967
Hinge pin for pump cover lever. . . . . . . . . . . . . . . . . . . 3450 5725
Exchange
Tools: Pin punch 1.8 mm, pin punch 6 mm, 4 mm gauge
1.
Open door and remove hinge pin with pin punch (1.8 mm)
from below. Do not lose torsion spring.
2.
Disassemble pump cover.
3.
Insert torsion spring in new pump cover. Insert hinge pin with
pin punch (6 mm) from above.
4.
Check pressure sensor with 4mm gauge in the Service Pro-
gram, and calibrate, if necessary.
Check
Safety check, pump unit check.
Fig.: 4 - 3
Hinge pin for
pump cover lever
Hinge pin for
pump cover
Torsion spring
Pump cover
Содержание 871 5416
Страница 1: ...Service Manual Infusomat fmS 0 Version 2 3 English 0 10 ...
Страница 10: ...Contact Persons 0 0 10 Infusomat fmS 2 3 gb For your notes ...
Страница 22: ...System Overview 1 1 12 Infusomat fmS 2 3 gb For your notes ...
Страница 30: ...Software 2 2 8 Infusomat fmS 2 3 gb For your notes ...
Страница 60: ...Unit Elements 4 4 20 Infusomat fmS 2 3 gb For your notes ...
Страница 64: ...Maintenance 6 6 2 Infusomat fmS 2 3 gb For your notes ...
Страница 78: ...Test Equipment and Special Tools 9 9 2 Infusomat fmS 2 3 gb For your notes ...
Страница 82: ...Spare Parts List 10 10 4 Infusomat fmS 2 3 gb For your notes ...
Страница 84: ...Index 11 11 2 Infusomat fmS 2 3 gb ...
Страница 86: ...Revision Documentation A A 2 Infusomat fmS 2 3 gb For your notes ...