Infusomat® fmS 2.3 gb
8 - 1
8
8 -
Procedural Instructions on the TSC
Visual Inspection
1.
Check the Infusomat® fmS and accessories for cleanliness.
2.
Check the Infusomat® fmS and accessories for completeness
and check configuration. Pay special attention to the follow-
ing parts:
a)
Check the special additional labels of the Infusomat®
fmS adjusted for the Infusomat® Space Line, please see
3.
Check the Infusomat® fmS and its accessories for damage
and the labels for readability. Pay special attention to the fol-
lowing parts:
a)
Pump sealing diaphragm
b)
Membrane keyboard
c)
Rubber feet
d)
Operating unit
e)
Lock mechanism of the pump cover
f)
Flow inhibitor
g)
Mains plug connector
h)
MFC plug connector
i)
Check fuse values
A fuse with T 0.315 A must be installed for devices for
100/110/120 V.
A fuse with T 0.16 A must be installed for devices for
200/230/240 V.
Fig.: 8 - 1
8 - 8
Содержание 871 5416
Страница 1: ...Service Manual Infusomat fmS 0 Version 2 3 English 0 10 ...
Страница 10: ...Contact Persons 0 0 10 Infusomat fmS 2 3 gb For your notes ...
Страница 22: ...System Overview 1 1 12 Infusomat fmS 2 3 gb For your notes ...
Страница 30: ...Software 2 2 8 Infusomat fmS 2 3 gb For your notes ...
Страница 60: ...Unit Elements 4 4 20 Infusomat fmS 2 3 gb For your notes ...
Страница 64: ...Maintenance 6 6 2 Infusomat fmS 2 3 gb For your notes ...
Страница 78: ...Test Equipment and Special Tools 9 9 2 Infusomat fmS 2 3 gb For your notes ...
Страница 82: ...Spare Parts List 10 10 4 Infusomat fmS 2 3 gb For your notes ...
Страница 84: ...Index 11 11 2 Infusomat fmS 2 3 gb ...
Страница 86: ...Revision Documentation A A 2 Infusomat fmS 2 3 gb For your notes ...