Service Program
3
3 - 8
Infusomat® fmS 2.3 gb
Menu
Function 450.0
The availability of menus on the user interface can be set.
-
Double rate entry
-
Online rate entry
-
Interval Bolus
1.
OK activates the function.
2.
Activate / deactivate the decimal function with NEXT.
3.
Acknowledge with YES.
4.
Return to main menu with END.
Delivery Rate
Function 460.0
The maximum and minimum delivery rates can be set. Range of
adjustment of the delivery rate: 0.1 to 999.9 ml/h
1.
OK activates the function.
2.
Select min./max. delivery rate with NEXT.
3.
Acknowledge with YES.
4.
Enter value with a numeric key.
5.
Acknowledge with OK.
6.
Return to initial function with END.
Air Alarm
Function 470.0
The air inline sensor sensitivity of the air rate alarm in ml/h (total
air alarm) and of the maximum air bubble in ml (single bubble)
can be adjusted.
Setting range air rate: 0.5 to 3.5 ml/h
Setting range air bubble: 0.01 to 0.3 ml/h
1.
OK activates the function.
2.
Select air rate (ml/h) or air bubble (ml) with NEXT.
3.
Acknowledge with YES.
4.
Enter value with a numeric key.
5.
Acknowledge with OK.
6.
Return to initial function with END.
Содержание 871 5416
Страница 1: ...Service Manual Infusomat fmS 0 Version 2 3 English 0 10 ...
Страница 10: ...Contact Persons 0 0 10 Infusomat fmS 2 3 gb For your notes ...
Страница 22: ...System Overview 1 1 12 Infusomat fmS 2 3 gb For your notes ...
Страница 30: ...Software 2 2 8 Infusomat fmS 2 3 gb For your notes ...
Страница 60: ...Unit Elements 4 4 20 Infusomat fmS 2 3 gb For your notes ...
Страница 64: ...Maintenance 6 6 2 Infusomat fmS 2 3 gb For your notes ...
Страница 78: ...Test Equipment and Special Tools 9 9 2 Infusomat fmS 2 3 gb For your notes ...
Страница 82: ...Spare Parts List 10 10 4 Infusomat fmS 2 3 gb For your notes ...
Страница 84: ...Index 11 11 2 Infusomat fmS 2 3 gb ...
Страница 86: ...Revision Documentation A A 2 Infusomat fmS 2 3 gb For your notes ...