Unit Elements
4
4 - 16
Infusomat® fmS 2.3 gb
i)
Remove weight and holder and do
not
save data.
j)
Call in calibration of pressure stage (function 540) in the
Service Program.
k)
Enter the calibration values via the keyboard and ac-
knowledge with ”yes”.
l)
Quit Service Program and save data.
m) Register the changed pressure values in the unit book.
n)
Check the electronic occlusion pressure
.
Check
Safety check, functional check.