Service Program
3
3 - 10
Infusomat® fmS 2.3 gb
Air Inline Sensor
Function 520.0
Alignment or check of the air inline sensor value (alarm threshold)
1.
OK activates the function.
2.
Press OK again to activate the air inline sensor value.
3.
The value can be changed with the entry keyboard.
4.
Acknowledge with YES.
5.
AIR SENSOR IS SET acknowledges the entry.
6.
Return to initial function with END.
7.
Quit the Service Program and save changes with YES.
Occlusion Level
Function 540.0
.
Pump Data
Function 550.0
Note
Compatibility between controller board and pump must be
checked before acknowledgement. Only press YES if pump corre-
sponds to controller board.
History
Function 560.0 *
The history function can be activated or deactivated.
1.
Press OK key twice to activate the function.
2.
Select the history function with NEXT.
3.
ON activates the function. The history protocol memory is in-
itialized (the current software version and the serial number
are registered).
Press OFF key to deactivate the function. The history protocol
memory is deleted.
4.
Return to initial function with END.
Note
See instructions for use for detailed information.
* Only software IFME
Содержание 871 5416
Страница 1: ...Service Manual Infusomat fmS 0 Version 2 3 English 0 10 ...
Страница 10: ...Contact Persons 0 0 10 Infusomat fmS 2 3 gb For your notes ...
Страница 22: ...System Overview 1 1 12 Infusomat fmS 2 3 gb For your notes ...
Страница 30: ...Software 2 2 8 Infusomat fmS 2 3 gb For your notes ...
Страница 60: ...Unit Elements 4 4 20 Infusomat fmS 2 3 gb For your notes ...
Страница 64: ...Maintenance 6 6 2 Infusomat fmS 2 3 gb For your notes ...
Страница 78: ...Test Equipment and Special Tools 9 9 2 Infusomat fmS 2 3 gb For your notes ...
Страница 82: ...Spare Parts List 10 10 4 Infusomat fmS 2 3 gb For your notes ...
Страница 84: ...Index 11 11 2 Infusomat fmS 2 3 gb ...
Страница 86: ...Revision Documentation A A 2 Infusomat fmS 2 3 gb For your notes ...