Procedural Instructions on the TSC
8
8 - 4
Infusomat® fmS 2.3 gb
Test Setup
Perform test setup with the subassemblies listed below, please see
also
An electronic manometer should be used for the measurement
described hereafter.
If a mechanic manometer is used instead values which are approx.
100 mbar lower are to be expected. Proceed as follows:
Note
The liquid level in the container must be approx. 80 cm above the
lower edge of the device.
-
Original Infusomat® Line
(new, can be used for the complete TSC incl. functional check)
(1 piece)
or
-
Infusomat® Space Line
(new, can be used for the complete TSC incl. functional check)
(1 piece)
-
Infusion bag or bottle, min. 500 ml
(1 piece)
-
Three-way valve
(2 pieces for measurement with electronic manometer)
-
10 ml syringe (air buffer for measurement with electronic
manometer)
(syringe drawn up with air to 10 ml and piston fixed mechan-
ically)
(1 piece)
-
1 ml syringe for air bubble injection
(1 piece)
-
Electronic manometer with peak value recognition
(1 piece)
-
Pressure cell (option of electronic manometer)
Diameter 100 mm / 160 mm, measuring range 2.5 bar / 4 bar
(1 piece)
-
Graduated cylinder 25 ml, +/- 0.4 ml
(1 piece)
Fig.: 8 - 3
1
2
10 ml air
Содержание 871 5416
Страница 1: ...Service Manual Infusomat fmS 0 Version 2 3 English 0 10 ...
Страница 10: ...Contact Persons 0 0 10 Infusomat fmS 2 3 gb For your notes ...
Страница 22: ...System Overview 1 1 12 Infusomat fmS 2 3 gb For your notes ...
Страница 30: ...Software 2 2 8 Infusomat fmS 2 3 gb For your notes ...
Страница 60: ...Unit Elements 4 4 20 Infusomat fmS 2 3 gb For your notes ...
Страница 64: ...Maintenance 6 6 2 Infusomat fmS 2 3 gb For your notes ...
Страница 78: ...Test Equipment and Special Tools 9 9 2 Infusomat fmS 2 3 gb For your notes ...
Страница 82: ...Spare Parts List 10 10 4 Infusomat fmS 2 3 gb For your notes ...
Страница 84: ...Index 11 11 2 Infusomat fmS 2 3 gb ...
Страница 86: ...Revision Documentation A A 2 Infusomat fmS 2 3 gb For your notes ...