Operation using the data setting device
Azbil Corporation
5-44
Model MCB10A
5-7-1 :
Configuration of OUTPUT CHECK MODE
Introduction
OUTPUT CHECK MODE has the following setting and adjustment items.
For details on the function of items, see the following pages.
Item
Content
Screen
OUTPUT
CHECK
I.OUT
Outputs a fixed value of analog
current output to perform loop
checks.
12.3 %
> OUTPUT CHECK
I.OUT
100%
OUTPUT
CHECK
P.OUT
Outputs a fixed value of pulse
output to perform loop checks.
12.3 %
> OUTPUT CHECK
P.OUT
100%
OUTPUT
CHECK
ST.OUT
Outputs a fixed value of contact
output to perform loop checks.
12.3 %
> OUTPUT CHECK
ST.OUT
CLOSE
LCD display flow
The screen flow of OUTPUTCHECK MODE is as follows:
> OUTPUT CHECK
I.OUT
100%
> OUTPUT CHECK
P.OUT
100%
> OUTPUT CHECK
ST.OUT
CLOSE
> OUTPUT CHECK
MODE
OFF