Operation using the data setting device
Azbil Corporation
5-18
Model MCB10A
Set the moving average processing in accordance with the following procedure:
Step
Procedure
Screen
1
The screen at right is a screen display
example in MEASURING MODE
(measurement state).
Press the MODE key.
20.0 %
01.94 m
3
/h
00069401
2
The OPERATOR'S MODE screen appears
for approx. two seconds and then the damping
setting screen appears.
20.0 %
* OPERATOR'S
MODE
20.0 %
* DAMPING
005.0 s
3
Press the
h
or
i
key to display the screen
shown on the right.
20.0 %
* AVERAGING
OFF
4
Press the
g
key to move the cursor to the OFF
position.
20.0 %
* AVERAGING
OFF
5
Press the
h
key to switch the OFF screen to
the ON screen.
Press the
g
key to move the cursor to the
value to be changed.
20.0 %
* AVERAGING
ON
01.0 s
6
Press the
h
or
i
key to display a value to be
set.
20.0 %
* AVERAGING
ON
05.0 s
7
Press the
g
key to move the cursor back to
the position under *.
Press the MODE key to return to the
MEASURING MODE and save data.
20.0 %
* AVERAGING
ON
05.0 s
CAUTION
You have only ten minutes to return to MEASURING MODE to save the new
value before the system resets it to the previously saved value.