Operation using the data setting device
Model MCB10A 5-15
5-5-4 : Resetting built-in counter totalized value
Reset the built-in counter to start totalization from a value set as the built-in counter
reset value. If this value is set to 1000, the built-in counter starts totalization from
1000 after the counter is completely reset.
If the built-in counter is reset, the built-in counter value just before the reset appears
at the side of PREV on the LCD display.
Set range: None
Default:
None
Reset the internal counter in accordance with the following procedure:
Step
Procedure
Screen
1
The screen at the right is a screen display
example in MEASURING MODE
(measurement state).
Press the MODE key.
20.0 %
01.94 m
3
/h
00069401
2
The OPERATOR'S MODE screen appears
for approx. two seconds and then the damping
setting screen appears.
20.0 %
* OPERATOR'S
MODE
20.0 %
* DAMPING
005.0 s
3
Press the
h
key three times to display the
screen shown on the right.
20.0 %
* CNT-RESET READY
00000000
PREV
4
Press the
g
key to move the cursor to the
position READY.
20.0 %
* CNT-RESET READY
00000000
PREV
5
Press the
h
key to reset.
In 0.5 seconds, the reset is completed with ON
changing to READY.
20.0 %
* CNT-RESET ON
00000000
PREV
20.0 %
* CNT-RESET READY
00123456
PREV
6
Press the
g
key to move the cursor to the
position under *.
Press the MODE key to return to the
MEASURING MODE and save data.
20.0 %
* CNT-RESET READY
00123456
PREV
CAUTION
You have only ten minutes to return to MEASURING MODE to save the new
value before the system resets it to the previously saved value.