Operation using the data setting device
Model MCB10A 5-27
5-6-3 : Setting detector information
Set detector information necessary for combination with the converter.
EX value:
Each detector has a unique calibration factor (EX value).
This value is determined at shipment in accordance with the actual
flow rate calibration. DO NOT change this value or the flowmeter
output will be incorrect.
Detector type:
Detector type is set as MCB.
Bore diameter: Sets the bore diameter (inside diameter) of the detector. The
correct bore diameter is set as factory default setting.
Setting range:
Detector constant: 200.0 to 699.9
Detector type: MCB
Bore diameter: 15 to 100
Set the detector information in accordance with the following procedure:
Step
Procedure
Screen
1
Enter ENGINEERING MODE in accordance
with the entry into ENGINEERING MODE
(see section 5-4-10). Then press the
h
or
i
key to display the screen at right.
12.3 %
#
MCB DIA 050.0
EX 300.0
2
Press the
g
key to set the detector constant.
Using the
h
or
i
key, enter the numeric value
found in the EX column stamped on the
nameplate of the detector to be combined.
12.3 %
#
MCB DIA 050.0
EX 320.0
3
Verify the detector type is MCB.
12.3 %
#
MCB DIA 050.0
EX 320.0
4
Then press the
g
key to select the bore
diameter.
Verify the bore diameter matches with the
diameter on the name plate.
12.3 %
#
MCB DIA 100.0
EX 320.0
5
Using the
g
key, move the cursor to the
position under #.
Press the MODE key to return to the
MEASURING MODE and save data.
12.3 %
#
MCB DIA 100.0
EX 320.0
CAUTION
You have only ten minutes to return to MEASURING MODE to save the new
value before the system resets it to the previously saved value.