2 Menu Item Selection
Reflex HC Series
57
specifications for communication of the balance to the default
settings, which are
[76],
[77],
[83],
[89],
[92], and
[94].
2.11.2 Handshaking
-
Handshaking determines whether the peripheral equipment
can receive communication data from the balance. This
function does not relay the status of the balance to the
peripheral equipment.
The balance is able to receive as long as there is space in the
receiving buffer of the balance. This function operates once
is displayed, operation in other states is not guaranteed.
When the balance output is retained by handshaking, the
display of the balance is locked. Determine the specifications
for handshaking.
Menu Map item [73]
Software handshaking
is not performed.
Menu Map item [74]
Software handshaking
is performed.
After the balance receives X-OFF (13H), the balance output is
retained.
After the balance receives X-ON (11H), the balance output is
initiated.
Menu Map item [75]
Hardware handshaking
is performed.
When DTR is OFF, the output from the balance is retained.
When DTR is ON, the output from the balance is initiated.
Menu Map item [76]
Timed hardware
handshaking is performed.
2.11.3 Format
Set the balance output format.
Menu Map item [77]
The standard format for the
Avery Weigh-Tronix electronic balance.
Note:
In this format, the number of the lowest place of Menu
Map item [70]
is assigned to identify the balance.
Available commands, functions and responses are limited.