2 Menu Item Selection
Reflex HC Series
47
Notes:
y
Use
the
key to erase the tare or zero the
balance at any time.
y
Pressing
the key while in the interval timer
standby state causes the power supply standby state to
be initiated.
y
To release the interval timer function, keep pressing the
key. This does not reset the interval timer function
to zero.
y
Using the interval timer function to record data over a
long period may cause data error due to balance drift.
y
Some instruments receiving the data may not operate
normally if the set time interval is short. To correct this,
set the time interval to a longer period. When the set
time interval is short and the instrument connected to
DATA I/O is unknown, it is recommended to set the
handshake to a setting other than
[76].
2.8.7
Auto-Memory and Zeroing
Use Menu Map item [51].
(1)
This menu is used to weigh a large number of individual
samples. The
(Auto-Memory and Zeroing) symbol is
illuminated when this function is active.
Load the weighing vessel and press the
key in
the Auto-Memory and Zeroing standby state (
and
symbols are lit).
Zeroing occurs.
(2) Press
the key.
The
symbol disappears, and Auto-Memory
and Zeroing measurement starts.
(3)
Load the first sample. Each time the stability symbol is
lit and the display is a value five times or more the Zero
Range or the
key is pressed, the displayed value is
output and zeroing occurs.