1 Basic Operation
Reflex HC Series
14
standby mode.
This symbol also illuminates when the
Application Measurement function has entered
the standby mode.
•
Inverse triangle
symbol
Illuminates when the solid specific gravity unit
is used. This symbol is also used as a
substitute for the decimal point.
*1
Stability symbol
The displayed value may change while the stability symbol
remains illuminated if the load is changing slowly or if the
stability detection band has been set to a large value.
1.2 Installation
1.2.1
Choosing the Installation Site
(1) Power
supply
CAUTION
•
Select an installation site that is near a power source to
ensure that the attached AC adapter is used properly.
•
Verify that the supply power voltage conforms to that
indicated on the AC adapter.
(2) Installation
site
CAUTION
Choose an installation site where the balance will be
protected from the following:
•
corrosive or flammable gasses.
•
dust, wind, vibration, electromagnetic waves, or
magnetic fields.
•
direct
sunlight.
•
extreme temperature or humidity.
•
rain or the possibility of being splashed with water.
Large capacity balances should be installed on a sturdy
floor and table that can withstand the total load of the
balance AND object to be weighed.