E1070 Service Manual
79
3.4 Service Menu
21.
Press the
UNITS
key…
CELLS
is displayed. Use this to check the mV/V level of each cell.
22.
Press the
key…
1
is displayed. This stands for cell 1.
23.
Press the
key…
on
or
oFF
is shown for the chose cell.
24.
Toggle between the choices by using the
TARE
or
UNITS
key and press
F1
to
select the displayed choice…
1
is displayed.
25.
Press the
UNITS
key to scroll to the next cell.
26.
Repeat steps
through
for each cell you need to enable.
27.
Repeatedly press the
SELECT
key until…
SEnSor
is displayed.
NETS (Network setup)
1.
With
SEnSor
displayed, press the
UNITS
key…
nEtS
is displayed. Use this to set up the networks that are used with the
indicator.
2.
Press the
key…
nEt 1
is displayed.
Cells must be enabled in order (1-4). Only one scale can be connected to the
SensorComm box.
Содержание E1070
Страница 1: ...Model E1070 Indicator Service Manual NORTH AMERICA AWT35 500106 Issue AF March 2010 ...
Страница 6: ...6 E1070 Service Manual ...
Страница 51: ...E1070 Service Manual 51 3 4 Service Menu Table 3 1 Printalbe Characters Chart ...
Страница 146: ...146 E1070 Service Manual 8 Network Connections ...
Страница 149: ...E1070 Service Manual 149 10 Complete Menu Structures 10 1 Complete Supervisor and User Menus Supervisor Menu User Menu ...
Страница 150: ...150 E1070 Service Manual 10 2 Complete Service Menu ...
Страница 153: ...E1070 Service Manual 153 11 3 Wall Mount Application ...
Страница 154: ...154 E1070 Service Manual 11 4 System Block Diagram ...
Страница 155: ...E1070 Service Manual 155 11 5 Keypad Schematic and Power Supply Bd P N 60071 0016 ...
Страница 158: ...158 E1070 Service Manual 11 8 I O External Calbe Identification Pin outs ...
Страница 162: ...162 E1070 Service Manual ...
Страница 163: ......