108
E1070 Service Manual
3
Using the Menus
Inbound Values
The following tokens may be affected by division size:
l
Tare
(token 2)
l
PLU Piece Weight
(token 5)
l
PLU Tare Value
(token 11)
l
PLU Lower Target Weight
(token 13)
l
PLU Upper Target Weight
(token 14)
If the division size of the scale is less than 1, the value that is mapped will be scaled
down by 10X. Where X equals the number of digits to the right of the decimal point.
Any floating-point values will not be scaled down.
Example
:
Token = 2
(Tare)
Type = 4
(Signed Long)
Division Size = 0.01
There are two digits to the right of the decimal point so the tare weight coming into the
indicator will be divided by 102 (100). If the desired tare weight is 99.56, a value of
9956 must be sent.
Error (Error Annunciator)
Follow these steps to configure the Network/Sensorcomm status annunciator.
1.
With
Error
displayed, press the
key…
The current setting will be displayed. This may be
oFF
,
S-Com
,
nEt 1
,
or
nEt 2
.
2.
Use the
UNITS
and
TARE
keys to change the setting, and press the
ENTER
key to accept the desired setting once it is displayed.
oFF
The annunciator will always remain off.
S-Com
The annunciator will show the status of the Sensorcomm
scale.
Red – a cell has been ghosted. Check the ghost log.
Green – a sensorcomm error has occurred. Check the
error log.
Off – Scale is functioning normally.
nEt 1
The annunciator will show the status of network #1.
nEt 2
The annunciator will show the status of network #2.
Red – A network error has occurred. Check the network
settings on the indicator and PLC, and reboot
the indicator.
Green – The network connection has been established.
Amber – The network is ready for a connection, but no
connection has been established.
This completes the Service menu. Press the
ESC
key to return to normal weighing
mode. You will be prompted to save the changes you’ve made. Press the
ESC
key to
abort any changes made and return to normal operating mode or press
ENTER
to
accept them and return to normal operating mode.
buSY
flashes on the display while
the unit saves data.
Содержание E1070
Страница 1: ...Model E1070 Indicator Service Manual NORTH AMERICA AWT35 500106 Issue AF March 2010 ...
Страница 6: ...6 E1070 Service Manual ...
Страница 51: ...E1070 Service Manual 51 3 4 Service Menu Table 3 1 Printalbe Characters Chart ...
Страница 146: ...146 E1070 Service Manual 8 Network Connections ...
Страница 149: ...E1070 Service Manual 149 10 Complete Menu Structures 10 1 Complete Supervisor and User Menus Supervisor Menu User Menu ...
Страница 150: ...150 E1070 Service Manual 10 2 Complete Service Menu ...
Страница 153: ...E1070 Service Manual 153 11 3 Wall Mount Application ...
Страница 154: ...154 E1070 Service Manual 11 4 System Block Diagram ...
Страница 155: ...E1070 Service Manual 155 11 5 Keypad Schematic and Power Supply Bd P N 60071 0016 ...
Страница 158: ...158 E1070 Service Manual 11 8 I O External Calbe Identification Pin outs ...
Страница 162: ...162 E1070 Service Manual ...
Страница 163: ......