E1070 Service Manual
141
6.3 Communications Timeout:
Transmit 20000 lb of Gross weight with Gross, Motion, and all Under bar(UB1-6)
segments turned on. A typical serial string is as follows:
Mode 3&4 Serial String:
G<SP>20000<SP>lb<CR><AN1><AN2><AN3><AN4><AN5><AN6><AN7><AN8><
CR>
<LF><AN1> = 0x00 <AN2> = 0x38 <AN3> = 0x00 <AN4> = 0x00
<AN5> = 0x01 <AN6> = 0x00 <AN7> = 0x3E <AN8> = 0x00
6.3
Communications Timeout:
If communications is lost between the host and the remote, the remote will display “-“,
middle dashes until a signal is acquired again. The timeout for this error to occur should
be 5-10 seconds in duration.
Table 3: Remote keypad press protocol
Command
Description of command
Z<LF>
Emulates a Zero key press
S<LF>
Emulates a Select key press
T<LF>
Emulates a Tare key press
P<LF>
Emulates a Print key press
U<LF>
Emulates a Units key press
F<LF>
Emulates an F1 key press
Содержание E1070
Страница 1: ...Model E1070 Indicator Service Manual NORTH AMERICA AWT35 500106 Issue AF March 2010 ...
Страница 6: ...6 E1070 Service Manual ...
Страница 51: ...E1070 Service Manual 51 3 4 Service Menu Table 3 1 Printalbe Characters Chart ...
Страница 146: ...146 E1070 Service Manual 8 Network Connections ...
Страница 149: ...E1070 Service Manual 149 10 Complete Menu Structures 10 1 Complete Supervisor and User Menus Supervisor Menu User Menu ...
Страница 150: ...150 E1070 Service Manual 10 2 Complete Service Menu ...
Страница 153: ...E1070 Service Manual 153 11 3 Wall Mount Application ...
Страница 154: ...154 E1070 Service Manual 11 4 System Block Diagram ...
Страница 155: ...E1070 Service Manual 155 11 5 Keypad Schematic and Power Supply Bd P N 60071 0016 ...
Страница 158: ...158 E1070 Service Manual 11 8 I O External Calbe Identification Pin outs ...
Страница 162: ...162 E1070 Service Manual ...
Страница 163: ......