AWP-TR1D-T3P
___________________________________________
Measured values:
Temperature and relative humidity from an external probe Digi/E
and up to three temperature from external probes Pt1000/C.
Other humidity quantities are calculated from the temperature
and humidity measured by the Digi/E probe.
Digi/E probe range, accuracy, and response time:
See manual for a particular Digi/E probe
Digi/E probe connection:
Digi/E probe is connected by 4-pin M8 ELKA 4008V connector.
Connection of pins is described at
Maximum length of cable for Digi/E probe cannot exceed 15 m.
Pt1000/C probe(s) range:
-90 °C to +260 °C (sensor Pt1000/3850 ppm)
-150 °C to +600 °C (for option AWP-TR1D-T3PF)
Measurement current ~250uA
Pt1000/C input accuracy (without probes):
±0.2 °C the at range below +100 °C
±0.2 % from measured value at range above +100 °C
Overall measurement accuracy of device with connected probe
is composed from accuracy of input and accuracy of used probe.
Pt1000/C probe(s) response time:
According used probe
Pt1000/C probe connection:
Two wire connection with capability for deviation compensation
due to resistance of uses cable. Connection is done via CINCH
(RCA) connector. Connection is described at
Recommended length of probe Pt1000/C is up to 15 m. Length
of probe cannot exceed 30 m. It is strongly recommended to use
shielded cable of probe.
Resolution:
Temperature including dew point temperature: 0.1 °C