Parameters of inputs
AWP-T
___________________________________________
Measured values:
Temperature from a probe which is part of shipment
Range:
-30 °C to +60 °C
Accuracy:
±0.4 °C
Response time:
t90 < 1 min (temperature step 20 °C, air flow 1 m/s)
Resolution:
0.1 °C (ADC resolution 16 bit)
Recommended calibration interval:
2 years
AWP-T1P
___________________________________________
Measured values:
Temperature from external probe ATAL Pt1000/C
Range:
-90 °C to +260 °C (sensor Pt1000/3850 ppm)
-150 °C to +600 °C (for option AWP-T1PF)
Measurement current ~250 uA
Input accuracy (without probes):
±0.2 °C the at range below +100 °C
±0.2 % from measured value at range above +100 °C
Overall measurement accuracy of device with connected probe
is composed from accuracy of input and accuracy of used probe.
Probe connection:
Two wire connection with capability for deviation compensation
due to resistance of uses cable. Connection is done via CINCH
(RCA) connector. Connection is described at
Recommended length of probe Pt1000/C is up to 15 m. Length
of probe cannot exceed 30 m. It is strongly recommended to use
shielded cable of probe only.