57
Error 2
A/D converter for measurement from Pt1000 probes is above
high limit. It is likely that temperature probe is not connected,
or cable is damaged.
Please inspect probe for a damage or connect probe.
Error 3
Measured value is outside expected range.
Please contact technical support.
Error 4
The source value for computed value (dew point) is not
available.
Please check relevant Digi probe if is properly connected and
is not damaged.
Error 10
Communication error with internal CO
2
sensor.
Please contact technical support.
Error 11
Measurement error from internal CO
2
sensor. One of reasons
for this error is an insufficient voltage from power source.
Please make sure that is used proper power supply adapter,
USB cable is not too long or damaged. In case pf issue will not
be resolved by using another adapter and cable, contact
technical support.
Error 15
Communication error with relative humidity sensor inside Digi
probe. It is likely that relative humidity sensor is damaged.
Please restart device by disconnecting from power source. If
error state persists, replace Digi probe.
Error 16
Measurement error from relative humidity sensor.
Please restart device by disconnecting from power source. If
error state persists, replace Digi probe.
Error 20
Unable to read calibration constants from internal barometric
pressure sensor.
Please contact technical support.
Error 21
Measurement error at internal barometric pressure sensor.
Please contact technical support.
Error 30
Communication error with internal A/D converter.
Please contact technical support.
Error 35 or n/a
Measured value from Digi probe is not available. It is likely that
Digi probe is not connected.
Please connect Digi probe and detect Digi probe(s).
Error 36
During Digi probe detection procedure was returned CRC error
of memory for calibration data.
Please contact technical support. They will provide software
WifiSensorUtility which will allow to restore content of memory
for calibration data from backup.