11
RECOVERABLE
errors do not stop the controller from attempting to complete a commanded move.
Large numbers of recoverable errors should be taken as a warning. Frequent servo errors (numbers 90-
92) often mean that the speed is near or exceeding the stage maximum. Frequent overtime errors
(numbers 1-9) often mean that competing processes, such as over-frequent serial status requests, are
using too much CPU time.
Change log
5/20/13
Branched from MS2000 manual
Vik
5/21/13
TG-1000 Prog started guide, integrated into this doc
Vik
6/20/2013
New options to
CCA Z
Vik
7/2/2013
Major edits to commands to better represent TG-1000s expectations Vik
7/26/2013
Expanded on Tiger banner, and other minor edits
Vik
7/30/2013
Build command improved
Vik
3/18/2014
Adjusted
RDSTAT
cmd function desp a bit
Vik
4/4/2014
Added TG-1000 addresses table
Vik
6/26/2014
Elaborated on CCA X?
Vik
CLEANING & Maintenance:
Cleaning should be done with a 70% isopropyl alcohol solution. Bottom air vents should be
checked to ensure that the vents do not become clogged with dust.
Maintenance should only be done by ASI, or an ASI approved technician.