![ASI Tiger TG-1000 Скачать руководство пользователя страница 34](http://html1.mh-extra.com/html/asi/tiger-tg-1000/tiger-tg-1000_instruction-manual_2982756034.webp)
11
Function:
This command returns the date and time the current firmware were compiled. It’s a
Card-Addressed command.
Example:
1CD
Dec 19 2008:16:19:59
This example shows that the firmware running was compiled on December 19th year
2008 at 4:19:59 PM.
Command: CNTS
Shortcut:
C
Format:
CNTS [axis] = [encoder counts per mm]…
Function:
Changes axis' encoder counts per mm. For example, doubling this number would cause
a given number of mm to be converted internally to twice as many encoder counts as
before. A command to move the stage 2 mm would instead cause it to move 4 mm. It’s
an
Axis-Specific Command.
MOST USERS DO NOT NEED THIS FUNCTION!
Example:
C X=13490.4
:A
c x?
:X=
13490.4
A
Changes the encoder constant on the X-axis to 13490.4 counts/mm. The default values
for this parameter are restored upon reset and should not require user modification.
Command: CUSTOMA
Shortcut:
CCA, CA
Format:
[Addr#]CCA X = ###
Function:
Configuration flags are set according to the table below for builds with
STNDRD_XY
and/or
STNDRD_Z
axes profiles. Configuration flags are changed one at a time for
each execution of the
CCA
command. The changes will not take effect until the
controller is restarted. Issue the
RESET
command to activate the new configuration.
It’s a
Card-Addressed command.
CCA X=
Description
Display Comment
5
XY Leadscrew Coarse Pitch (6.35 mm - Standard)
B
Firmware default
6
XY Leadscrew Fine Pitch (1.59 mm)
A
7
XY Leadscrew Super Coarse (12.7 mm)
C