11
Shortcut:
B
Format:
BACKLASH [axis] = [distance]…
Units:
millimeter
Type:
Axis-Specific
Function:
This command sets (or displays) the amount of distance in millimeters to travel to
absorb the backlash in the axis' gearing. This backlash value works with an anti-backlash
routine which ensures that the controller always approaches the final target from the
same direction. A value of zero (0) disables the anti-backlash algorithm for that axis;
zero is default.
Example:
B X=.05 Y=.05 Z=0
:A
B x?
:X=0.040000 A
The command in this example will make the controller move the X and Y axes to a
location 50 microns away from the final target before moving to the final target, while
the anti-backlash algorithm for the Z axis is disabled.
Command: BCUSTOM
(Version 3+)
Shortcut:
BCA