32
Caregiver obligations shall be carried out by personnel with
sufficient knowledge following the instructions in this man-
ual.
Every day
Clean/Disinfect.
•
Calypso
has to be cleaned immediately after every use.
ArjoHuntleigh cleaning agents and shampoos should be
used in
recommended concentrations labelled on the
disinfectant bottle
.
See further disinfecting/cleaning instructions
Every week
Visually check all exposed parts.
•
Examine Calypso for damage.
Visually check mechanical attachments.
•
Check the mechanical attachments by placing a foot on
the chassis and pull the seat upwards with both hands.
No gaps are allowed (
1
).
Visually check the safety strap/leg strap.
•
Check the complete length of the strap for fraying, cuts,
loose stitching. If the strap is found damaged in any way,
replace it
(2)
.
•
Visually check all exposed parts, especially where per-
sonal contact is made by either the resident or caregiver.
Make sure no cracks, sharp edges or unhygienic areas
that could cause the resident or user injury. Replace dam-
aged parts.
Check hand control and cable.
•
Visually check the condition of the hand control and its
cable. If found cut or damaged, replace. Electric version
(3)
.
Check and clean the castors.
•
Make sure the castors are properly fixed and are rolling
and swivelling freely. Clean with water. (the function can
be affected by soap, hair, dust and chemicals from floor
cleaning)
(4)
.
Check the cushion.
•
Make sure the cushion has no cracks or tears that would
allow water to get into the filling. If there is such damage,
replace damaged parts CDB6000
(5)
.
Visually check the battery and the battery charger.
•
Check the cable condition as well. Withdraw and replace
immediately if damaged. Electric version
(6)
.
Perform functionality test.
•
Test up/down motion by pressing up/down.
•
Test emergency High/Low
Yearly
Calypso
has to be serviced according to the Preventive
Maintenance Schedule (Qualified Personnel Action/Check).
Содержание CALYPSO
Страница 1: ...CALYPSO Instructions for Use 04 CD 03 7GB September 2009 0086 ...
Страница 8: ...8 ...
Страница 11: ...11 1 2 3 b a b a ...
Страница 13: ...13 1 2 3 4 A B A B ...
Страница 15: ...15 1 2 3 ...
Страница 17: ...17 2 3 4 5 1 CDB60XX CDB70XX ...
Страница 19: ...19 1 2 3 4 5 6 ...
Страница 21: ...21 1 2 3 ...
Страница 23: ...23 1 3 4 8 9 ...
Страница 25: ...25 3 4 6 10 MIX MIX MIX MI ...
Страница 27: ...27 a 1 2 ...
Страница 29: ...29 1 1 2 6 ...
Страница 33: ...33 1 2 3 4 5 6 ...
Страница 43: ...ArjoHuntleigh Hospital Equipment AB PO Box 61 Verkstadsvägen 5 SE 241 21 Eslöv Sweden ...