30
Trouble shooting
Trouble shooting- Calypso
Problem
Action
Using the lift
The lift continues to raise or descend after the hand
set button is released
Activate the red emergency stop button and pull the
battery out - remove the resident from the lift and call
for authorised service
The lift does not raise or descend when the hand set
button is pressed
1. Make sure that the battery is fully engaged in the
battery compartment
2. Make sure that the hand control connection plug is
properly engaged in the socket
3. Make sure that the battery is charged
Warning: Do not leave the resident unattended
while doing this!
4. Use the emergency lowering to descend the resi-
dent. call for authorised service as soon as the resi-
dent is removed from the lift.
The lift is hard to manoeuvre during transport.
1. Make sure all brakes are released
2. Make sure all castors roles and turns freely
- if not, remove the resident from the lift and call for
authorised service
The lift is making abnormal sounds during lifting or
transport
Remove the resident from the lift and call for author-
ised service
The lift beeps shortly once every minute when not
operated
Contact ArjoHuntleigh technical service personnel
Содержание CALYPSO
Страница 1: ...CALYPSO Instructions for Use 04 CD 03 7GB September 2009 0086 ...
Страница 8: ...8 ...
Страница 11: ...11 1 2 3 b a b a ...
Страница 13: ...13 1 2 3 4 A B A B ...
Страница 15: ...15 1 2 3 ...
Страница 17: ...17 2 3 4 5 1 CDB60XX CDB70XX ...
Страница 19: ...19 1 2 3 4 5 6 ...
Страница 21: ...21 1 2 3 ...
Страница 23: ...23 1 3 4 8 9 ...
Страница 25: ...25 3 4 6 10 MIX MIX MIX MI ...
Страница 27: ...27 a 1 2 ...
Страница 29: ...29 1 1 2 6 ...
Страница 33: ...33 1 2 3 4 5 6 ...
Страница 43: ...ArjoHuntleigh Hospital Equipment AB PO Box 61 Verkstadsvägen 5 SE 241 21 Eslöv Sweden ...