10 Isolation resistance measurement (Iso.)
ARGUS Copper Box
51
10 Isolation resistance measurement (Iso.)
Isolation resistance measurement can indicate damage to cable isolation, moisture
penetration or oxidised contact points on the line.
A voltage check is performed before the measurement in which the voltage
limits given in the table (see page 11) may not be exceeded.
See the chapter Technical data page 14 for information on measuring ranges, resolutions
and accuracies of isolation resistance measurement.
Isolation resistance measurement settings
ARGUS in the status display.
Isolation resistance measurement is not
yet started.
Before the measurement, the selected
test parameters are displayed.
- Measuring voltage (in this example 105 V)
- Measuring current (in this example
max. 2 mA)
- Polarity (in this example standard)
Hotkeys/settings of Isolation resistance
measurement s. page 20 and in the sheet
under.
The configuration options can vary depending on the selected sockets.
<Menu>
Opens the Copper Box
<Pol.
inverse>
The polarity is inverted/
reversed, see page 52.
<Start>
Starts the isolation
resistance measurement.