SH-7000-Chroma-EN-04/17
8
!!IMPORTANT!!:
To ensure that the basic settings take effect, set address 50 = 0!
Optional settings, Address 31 and 32, have to be programmed with
PremiumLink™ SW!
Ensure that after maintenance the configuration on address 50 = 1-8.
Read the values on pump screen.
2.3
Basic Settings
Address
Description
Default
0
9999: Re-initialisation (by replacing main PCB) thereafter it
is mandatory to recalibrate the pump.
0
1
9999: copy setup from display to mains.
0
2
9999: copy setup from mains to display.
0
3
Enter update codes. Step 1: Update ID 1: STD, 2: DCA, 3:
PCA, 4: TCI, 5: RAM, 6: TOC, 7: PIB, confirm with select.
Step2: send the upgrade code, update ID and displayed
code to Arcomed, do not switch off device until step 3 is
completed. Step 3: enter the upgrade password supplied
from Arcomed and confirm with select.
0
4 - 7
Reserved
0
8
Door Lock Offset (Not applied on 7000)
1300
9
LCD Contrast: allows adjusting the LCD contrast
(Not applied on Chroma)
354
10
Speed correction, see volume test
900
1100
11
Rate tuning, allows to fine tune the speed depending on the
rate. Internal setting.
0
12
Upstream pressure sensor trigger level, only available if an
upstream sensor is installed. If installed, normally set to 100,
otherwise set to 0 to disable.
0
13
Long time correction LTC, PVC = 2020,
Silicon = 0, deactivated. Allows to fine tune speed for long
term infusions. Internal setting.
2020
15
Gain of pressure transducer (see pressure calibration)
300
700
16
Offset of pressure transducer when door is closed with IV
set (see pressure calibration)
300
700
17
Limit pressure if door closed without IV set (see pressure
calibration)
20
400