SH-7000-Chroma-EN-04/17
38
11. Select address 1 in RATE display, enter the value 9999 in DATA display. In your
screen will appear
“d
m 9999”
confirm with “SELECT” button. Then your
calibration data will be transfer from MAIN pcb to DISPLAY pcb. It will take a
couple of seconds. When
“d
m 0”
the transfer is completed.
12. Turn the pump off and on again.
13. In order to ensure the proper operation of the pump, perform your operation and
safety tests.
14.
If it does success: go to 21.
If it doesn’t success and the pump goes to defect, some calibration data could be
corrupted or lost. Please go to 15.
15. Turn ON the pump in Setup Mode.
16.
Select address 0, enter the value 9999, confirm with “SELECT” button.
17.
Select address 51, enter the production date of the device, confirm with “SELECT”
button.
18.
Select address 52, enter the serial nr. of the device, confirm with “SELECT”
19. Go through all pump calibrations. Go to chapter 7.
20. In order to ensure the proper operation of the pump, perform your operation and
safety tests.
21. Update complete