SH-7000-Chroma-EN-04/17
6
measured and used for calculation of the pressure inside the system. The pressure sensor
requires a mechanical and electrical calibration (mbar).
The Easy Clip functions as a drop counter and is fastened to the drip chamber. Drops are
detected via light barrier in the Easy Clip. The Volumed®
µVP7000 Chroma is not
regulated through the drop count, the Easy Clip serves as an additional safety feature
allowing to detect reverse flow, free flow, upstream occlusion or and empty bag.
Using a hall sensor the rotation of the peristaltic system is monitored. If the measured
speed does not correspond to the set rate, the pump is stopped and an acoustic alarm is
emitted. This allows checking the mechanical and electrical functioning of the peristaltic
system.
All verification tests were performed with the IV sets Volumed
®
Set Standard APPF512
and the Volumed
®
Set silicon APSF227.
Software:
The microcontrollers are equipped with an internal flash memory (512kB). For software
updates and data transfer the Volumed®
µVP7000 Chroma is equipped with infrared
interfaces. This enables to carry out these updates without having to open the housing.
The required USB interface (Part No
: 5316) and arcomed ag’s own PremiumLink™
software are compatible with all common Window operating systems.
!!WARNING!!:
The Volumed®
µVP7000 Chroma is turned completely off during a software update to
access the internal flash memories. At this stage any false manipulation may cause pump
failures. It is of utmost importance that only trained personnel carry out updates because
there are several variations of carrying out updates. arcomed ag is not responsible for
pump damages because of false handling. There is no physical damage done to the
pumps, but the processors need to be reinitialized. Only arcomed ag may carry out the
task of re-initialisation. Furthermore are updates only to be carried out by trained
personnel.
Specifications, instructions for use, warranty, performance data, EMC behaviour:
The specifications, instructions for use, warranty conditions, performance data and EMC
behaviour can be found in the user manual.
Maintenance:
Maintenance and safety checks have to be done according to the checklist in the user
manual.
Worn or defective parts must be replaced. However, the pumps do not need any regular
replacement of specific items. Arcomed AG provides service kits and it might be more
efficient to use all parts provided in the kits to avoid additional repairs. It is however not
mandatory to use all parts in the kits.