Internal Alarms
The Volumed® µVP7000 Chroma continuously monitors the correct functionality of the
pump. For safety reasons the pump will trigger an internal alarm if any discrepancy is
detected. If an internal alarm occurs, the pump will automatically stop and emit a
permanent audible alarm and the alarm pictogram is illuminated. An internal alarm does
not mean that the pump is defect, but it should be carefully checked by the technical staff
before putting it back into operation.
HINT: Take the pump of the mains power. You can switch off the pump after an internal
alarm by pressing the ON/OFF key. If the pump should not power down immediately and
continue the audible alarm, hold the ON/OFF key of 12 seconds to reset it. A reset will not
erase the configuration of the pump. However the current infusion settings might be lost.
In the “Status / Help” menu, which can be accessed with the “SELECT” key, the LCD will
display the error code. In such cases the pump must not be used any more until the cause
has been checked. If the device cannot be repaired the pump should be sent to technical
department or to arcomed ag.
4.1
Meaning of error codes:
Code
Meaning
Cause
Measure
F-00
Communication
error between
processors
Different Software on
Display PCB and Main
PCB
Complete SW update
Main PCB defect
Replace Main PCB
Display PCB defect
Replace Display PCB
Display cable loose
Fix display cable
F-01
Air detector check
Air detector plate
broken
Replace air detector plate
Front PCB broken
Replace Front PCB
Display PCB broken
Replace Display PCB
F-03
Motor too slow (Hall
counter check)
Hall sensor defect.
Replace Hall PCB
PCB Motor driver
defect.
Replace Main PCB
Stepper motor defect
Replace stepper motor
Pump block not
greased well
Clean peristaltic mechanism and
reapply grease
Hall PCB not
connected properly
Check Hall PCB connector on Hall
PCB and Main PCB
Screw motor axis loose Tighten screw motor axis
Motor axis defect
Replace motor axis
Magnet in crown gear
lost
Replace magnet
F-05
Motor too slow
(Volume counter
check)
SW Problem
SW update necessary
Main PCB defect
Replace Main PCB