SH-7000-Chroma-EN-04/17
25
6.
Place new cover in Front Housing
7.
Apply a small amount of EP grease to the Pump Block fingers.
8.
Mount pump block and reassemble pump in reverse order.
Check calibration and install Main pcb and battery. Close the housing.
5.7
Replacement of Front Housing Air Detector Plate
Tools & recommended spare parts:
- Antistatic Work place
-
Philips screw driver No. 1
- 5.5 mm Hex driver
- 2.5 mm Hex driver
- 2 mm Hex driver
- Air in Line Detector Kit (76046)
Procedure:
1.
Dismantle Main PCB (5.1)
2.
Dismantle pump block (5.2)
3.
Dismantle Stop Flow Mechanism (5.5)