112
Technical Specifications and Operating Parameters
12.5
Antenna System
The included antenna is directly mounted on a 26 mm pole (1.05 in OD or 0.75 in ID pipe), with either a
standard 1 in – 14 (approximately M25.4
×
1.81) marine-mount thread or a 0.75 in NPT pipe thread. Other
mounting configurations are available (contact Arbiter Systems).
GNSS Antenna Assembly: 0.75 in pipe thread mount, 35 dB gain; 3.3 Vdc to 5 Vdc.
Optional antenna mounting bracket available to mount on 60 mm (2 in, nominal) pipe (See
section 4.1.2).
12.5.1
Antenna Cable
15 m (50 ft) cable included with antenna.
Other cable styles and lengths available, see Table 4.2 under Available Antenna Cables and Accessories
for Longer Runs.
12.6
Operator Interface
12.6.1
Setup Methods
Clock setup: Via RS-232C Interface, use supplied utility software
8 Front-panel keys
NTP/PTP Server, use Ethernet port web interface
12.6.2
Setup Functions
Serial COM1
configures port parameters; returns setup and broadcast ASCII string; transmits RS-
422/485.
Serial COM2
: configure port parameters, return setup and broadcast ASCII string.
Local Time & DST
: set local time zone offset from UTC, and daylight saving time offsets.
Out of Lock
: Sets the time before an out-of-lock condition alarms. Can be turned off. Counts up to
45 days and clears when clock regains GNSS synchronization. See Section 8.4.
Relay Config.
: The internal relay may be set to make/break for the following conditions: (1) out of
lock, (2) fault, (3) alarm, (4) clock stabilized, and (5) failsafe.
Back Light
: Sets the back light in the LCD display to be ON, OFF, or in Auto mode.
System Delays
: Offsets the time delay in nanoseconds for the GNSS signal to pass from the antenna
to the GNSS receiver.
Programmable Pulse
: Sets the mode, interval and pulse width for programmable pulses. Also has
an auxiliary IRIG-B mode, independent from the main IRIG-B setting.
IRIG Time Data
: Sets the time zone and IEEE C37.118.1 mode.
Event/Deviation
: Set to either Event or 1 PPS Deviation.
Option Control
: Configures the option board function in the clock, if installed.
C Display Format
: Configures the date format for the large display on Model 1201C clocks to
MM/DD/YY or DD.MM.YY.
12.6.3
Display
Model 1200B and Model 1201B/C include a 2-line by 20-character LED backlit LCD
Model 1201C also includes a six-character, 20mm LED time display
Содержание 1200B
Страница 4: ...iv ...
Страница 135: ...A 3 Physical Dimensions 117 Figure A 2 Suggested Mounting of the GNSS Surge Arrester ...
Страница 143: ...B 7 Four Fiber Optic Outputs 125 Figure B 4 Jumper Locations ...
Страница 145: ...B 8 8 Channel High Drive IRIG B Amplifier 127 Figure B 5 8 High Drive Outputs Jumper Locations ...
Страница 161: ...B 10 Four Additional Outputs and Dry Contacts 25 50 Vdc 143 Figure B 7 Option Connector Signal Locations ...
Страница 212: ...Appendix E Statement of Compliance The following page is a statement of compliance that includes Model 1201B and 1201C ...