68
10.7
CALLIBRATION
The conductivity requires periodic calibration. Calibration is usually required after cleaning the sensor. APRO con-
troller requires multipoint calibration: calibration solutions of 1413 mkS, 84 mkS, and air for 0 mkS calibration.
a.
Switch the system to OFF position.
b.
Take off the conductivity sensor. Wipe it with the distilled water and dry it.
c.
Prepare the calibration solutions of 1413 mkS and 84 mkS. Make sure they are not expired. It is recommended to
perform the calibration under 20-25
°C
(the room temperature).
d.
Hold
▲
button for 10 seconds. The main calibration menu will appear:
Make sure both of the conductivity sensors show 0 mkS
and represent the actual temperature. If the sensor’s
value is not close to 0, or the temperature sensor represents the wrong value, the corresponding sensor should be
replaced.
e.
Press
⬛
button. Permeate
conductivity ‘0 mkS’ calibration screen
will appear:
Make sure the conductivity sensor is in the air and press OK
⬛
button. Hold the sensor into the air while the
screen shows WAIT mode.
Содержание APRO 1000 LPH
Страница 1: ...aquaphor pro com TP APRO WH R02 APRO WH 1000 LPH WATER PURIFICATION SYSTEM...
Страница 8: ...8 aquaphor pro com TP APRO WH R03 2 SYSTEM OVERVIEW APRO WH 1000 LPH system dimensions...
Страница 9: ...9 aquaphor pro com TP APRO WH R03...
Страница 13: ...13 aquaphor pro com TP APRO WH R03 3 2 ELECTRIC CIRCUIT SCHEMES...
Страница 14: ...14 aquaphor pro com TP APRO WH R03...
Страница 15: ...15 aquaphor pro com TP APRO WH R03...
Страница 16: ...16 aquaphor pro com TP APRO WH R03...
Страница 17: ...17 aquaphor pro com TP APRO WH R03...
Страница 18: ...18 aquaphor pro com TP APRO WH R03...
Страница 19: ...19 aquaphor pro com TP APRO WH R03...
Страница 20: ...20 aquaphor pro com TP APRO WH R03...
Страница 80: ...80 14 ADDITIONAL MATERIALS...
Страница 81: ...81...
Страница 83: ...83 REMARKS...