Chapter 3 Scanning Operations
3-12
Applied Biosystems
3
Enables you to print out the currently displayed data or print
the file report using the format selected on the Export/Print
dialog box.
To only print the data displayed on the screen, hold down the
Shift key while clicking on the Print button.
To print the file report, click on the Print button.
Plate In/Plate Out
Causes the scanner to move the plate carrier into or out of the
scanner so you can insert or remove a plate.
•
To move the carrier out of the scanner, click on the
Plate Out button.
•
To move the carrier into the scanner, click on the Plate
In button.
Lamp On/Lamp
Off
Allows you to turn the lamp on or off:
•
To turn the lamp on, click on the Lamp On button.
•
To turn the lamp off, click on the Lamp Off button.
Scanning with the
lamp off
You can scan with the lamp off by selecting Lamp Off from the
Excitation menu on the Main Setup screen before you begin a
scan. Scanning with the lamp off allows you to measure
bioluminescence (glow). This allows you to use the CytoFluor
Series 4000 as a luminometer, although all luminescent
assays may not be optimized on the system.
Cycle
You use this field to indicate the cycle you want to display after
the system acquires data. To select the cycle:
1.
Click on the down arrow in the Cycle field to display the
pull-down menu.
2.
Select a value of 1 to 200 (or the last cycle acquired
during your run).
NOTE: This field displays the cycle of a run while
acquiring data.
Scan
Enables you to select the scan you want the application to
display of the currently active cycle. (The active cycle depends
on what you selected for the Cycle field.) When setting up a
run, use this box to select the scan parameter you are setting.
See Section 3.3, Setting Up Multiple Scans in a Cycle, for
details.