Chapter 6 Customizing the CytoFluor System
6-18
Applied Biosystems
6
When properly adjusted, the probe tip is 0.4 mm (the
thickness of the 0.015-inch height adjustment spacer)
away from the probe.
Adjusting probe
height in bottom-
read
1.
Loosen the two screws on the rear of the probe
mounting block. Do not remove them. Leave the probe
connected to the mounting bracket.
2.
Slide the probe all the way down.
3.
Slide the 0.015-inch spacer against the bottom of the
plate carrier and move the carrier over the probe head.
4.
Bring the probe head up to just touch the 0.015-inch
spacer. Do not place excess force on the plate carrier, or
you could damage it.
5.
Tighten the two screws on the rear of the probe
mounting block. Gently remove the 0.015-inch spacer.
6.
Slide the plate carrier to the back of the instrument.
7.
Make sure the fiber-optic cables are inside the unit.
CAUTION
The fiber-optic cables can be damaged if they are
caught between the cover and the base of the scanner.
8.
Close the scanner cover and seat it properly on the
gasket. Turn the thumbwheel counterclockwise until finger
tight. Turn on the scanner.
CAUTION
Do not over-tighten or under-tighten the cover thumbwheel
or the plate carrier cannot exit the scanner. To make sure
you closed the cover correctly, click Plate Out. If the plate
carrier does not move out, turn off the scanner and turn the
thumbwheel further to tighten it.