Defining Plates
CytoFluor User’s Guide
5-19
5
9.
Verify that the well is centered over the probe.
NOTE: If your scanner has a temperature control
chamber, gently lift up the edges of the slit in the
insulation on the temperature control chamber to see
the well.
If the well is not centered, center it by using the scroll
bar:
•
Drag the scroll box to move the plate in the
direction you drag the scroll box.
•
Click the bar to move the plate in increments
of 20 units in the direction of the arrow.
•
Click the arrow to move the plate in
increments of one unit in the direction of the
arrow.
10. Select X2-Y2.
The last well in the plate is positioned over the probe.
11.
Verify that the well is centered over the probe.
If the well is not centered, center it by using the scroll
bar as described in step 9.
12.
Click Test Move. The probe resets to the home position
and returns to the previous position.
13.
Click OK.
Click Cancel if you do not want to save the coordinates.