Replacing Optical Filters
CytoFluor User’s Guide
6-3
6
Figure 6-1 Filter Assemblies
1.
Make sure the plate transport carrier is inside the
scanner. If it is not inside, click on the Plate In button.
2.
On the main setup screen, set the Gain to 0 and turn the
lamp off by clicking the Lamp Off button. (Leave the
power on so you can use the Maintenance options.)
3.
Open the cover by turning the thumbwheel
counterclockwise and lifting up the cover.
Figure 6-2 Opening the Scanner
4.
Remove the two screws that secure the top cover of the
filter wheel assembly. Then remove the cover.
PB100555
Excitation filter
assembly
Emission filter
assembly
Thumbwheel