99
Modbus
(
RTU
)
Command
3101
Comparator mode
0000
:
ABS
0001
:
PER
0002
:
SEQ
Read and write registers, 2-byte
integer
3102
Secondary parameter comparison
on/off
0000:
Secondary
parameter comparison
off
0001
:
Secondary
parameter comparison
on
Read and write registers, 2-byte
integer
3103
Bin count
0001~0009
Bin count 1~9bin
Read and write registers, 2-byte
integer
3104
Beep
0000
:
off
0001
:
GD beep
0002
:
NG beep
Read and write registers, 2-byte
integer
310A
Primary parameter nominal value 4-byte floating point
number
Read and write registers, data
occupies 2 registers
310C
Secondary parameter lower limit
value
4-byte floating point
number
Read and write registers, data
occupies 2 registers
310E
Secondary parameter higher limit
value
4-byte floating point
number
Read and write registers, data
occupies 2 registers
3110
Primary parameter BIN1 lower
limit value
4-byte floating point
number
Read and write registers, data
occupies 2 registers
3112
Primary parameter BIN1 higher
limit value
4-byte floating point
number
Read and write registers, data
occupies 2 registers
3114
Primary parameter BIN2 lower
limit value
4-byte floating point
number
Read and write registers, data
occupies 2 registers
3116
Primary parameter BIN2 higher
limit value
4-byte floating point
number
Read and write registers, data
occupies 2 registers
3118
Primary parameter BIN3 lower
limit value
4-byte floating point
number
Read and write registers, data
occupies 2 registers
311A
Primary parameter BIN3 higher
limit value
4-byte floating point
number
Read and write registers, data
occupies 2 registers
311C
Primary parameter BIN4 lower
limit value
4-byte floating point
number
Read and write registers, data
occupies 2 registers
311E
Primary parameter BIN4 higher
limit value
4-byte floating point
number
Read and write registers, data
occupies 2 registers
3120
Primary parameter BIN5 lower
limit value
4-byte floating point
number
Read and write registers, data
occupies 2 registers
3122
Primary parameter BIN5 higher
limit value
4-byte floating point
number
Read and write registers, data
occupies 2 registers
3124
Primary parameter BIN6 lower
limit value
4-byte floating point
number
Read and write registers, data
occupies 2 registers
3126
Primary parameter BIN6 higher
limit value
4-byte floating point
number
Read and write registers, data
occupies 2 registers