107
Modbus
(
RTU
)
Command
0000
Internal trigger
0001
Manual trigger
Use Trigger key
0002
External trigger
Handler trigger
0003
Remote trigger
SCPI trigger
12.3.7
Measurement Frequency Register [3006H-3007H]
Write [1kHz: 1000 = 44 7A 00 00]
1
2
3
4
5
6
7
8~11
12
13
01
10
30
06
00
02
04
44 7A 00 00
12
AD
write
register
number of
registers
byte
data
CRC
Response:
1
2
3
4
5
6
7
8
01
10
30
06
00
02
AE
C9
register
number of
registers
CRC
Read
1
2
3
4
5
6
7
8
01
03
30
06
00
02
CF
1A
read
register
number of
registers
CRC
Response:
1
2
3
4~7
8
9
01
03
04
44 7A 00 00
CF
1A
byte
data
CRC
12.3.8
Level Register [3008H-3009H]
Write [1V: 1.00 = 3F 80 00 00]
1
2
3
4
5
6
7
8~11
12
13
01
10
30
08
00
02
04
3F 80 00 00
EB
B4
write
register
number of
registers
byte
data
CRC
Response:
1
2
3
4
5
6
7
8
01
10
30
08
00
02
FF
09
register
number of
registers
CRC
Read
1
2
3
4
5
6
7
8
01
03
30
08
00
02
7A
CA
read
register
number of
registers
CRC
Response:
1
2
3
4~7
8
9