75
SCPI Command Reference
Note
The suffix unit
Ω
can't be used with this command.
This command CANNOT be used in LIST SWEEP DISPLAY page and CORRECTION page.
10.9.4
LEVel:ALC (=AMPlitude:ALC)
The LEVel:ALC or AMPlitude:ALC command enables the Automatic Level Control (ALC).
Command Syntax
LEVel:ALC {on,1,off,0}
AMPlitude:ALC {on,1,off,0}
Parameter
{on,1,off,0}
Where,
on (1)
Enable ALC
off(0)
Turn off the ALC.
Example
SEND> LEV:ALC on
Query Syntax
LEV:ALC?
or AMP:ALC?
Query Response
{on,off}
Example
SEND> LEV:ALC?
RET>
off
Note
This command CANNOT be used in LIST MEAS page , CORRECTION page and DCR
mode.
10.10
APERture Subsystem
The
APERture
subsystem command sets the integration time of the ADC and the averaging
rate.
Figure 10-5
APERture Subsystem Command Tree
APERture
{SLOW,MED,FAST}
<averaging rate value:NR1>
Command Syntax
APERture {SLOW,MED,FAST}
APERture <value>
SPEED(spd) {SLOW,MED,FAST}
SPEED(spd) <value>
Parameter
Where,
SLOW
Set test speed to slow
MED
Set test speed to medium
FAST
Set test speed to fast
<value>
NR1(0 to 256): Averaging rate (0=OFF=1)
Example
SEND> APER FAST
SEND> APER 10
Query Syntax
APER?
Query Response
{SLOW,MED,FAST},<avg value>
Example
SEND> APER?
RET>
slow,0
10.10.1
APERture:RATE?
The
APERture:RATE?
query returns the current integration time.
Query Syntax
APER:RATE?
Query Response
SLOW