AT381x User's Guide
102
1
2
3
4
5
6
7
8
01
03
20
02
00
02
6E
0B
passive
station
read
register
number of
registers
check code
Response:
1
2
3
4
5
6
7
8
9
01
03
04
50
15
02
F9
3B
D5
01
03
byte
single precision floating point
number
CRC-16
Among them, B4~B6 is measurement data
:
501502F9 stands for 1E10
(
Low position in front
)
12.2.2
Fetch Comparator Results [2004H]
Register 2004 records voltage and resistance comparator results
16-bit storage domain:
Among them
:
BIT8
secondary parameter bin
1
:
NG
BIT7
total GD bin
1
:
primary and secondary parameter GD
,
0
:
Total NG
BIT3~BIT0
GD bin
1~9
:
GD
0
:
NG
Send:
1
2
3
4
5
6
7
8
01
03
20
04
00
01
CE
0B
passive
station
read
register
number of
registers
check code
Response:
1
2
3
4
5
6
7
01
03
02
00
01
E0
E5
Example
:
Data 01 81
BIT8:
1=secondary parameter NG
BIT7:
1=
Total NG
BIT3-BIT0
:
1=
primary parameter GD bin
BIN1
Data 00 01
BIT8
:
0=secondary parameter GD
BIT7
:
0=Total GD
BIT3-BIT0
:
1=
primary parameter GD bin
BIN1
Fetch primary and secondary measurement values and comparator results
【
2000~2004
】
Send:
1
2
3
4
5
6
7
8
01
03
20
00
00
05
8E
09
passive
station
read
register
number of
registers
check code
Response:
01 03 0A
44 79 D4 B1
37 D6 9D C2
00 81
C6 24