AT381x User's Guide
48
STEP
Manual Trigger. Each time the instrument is triggered by [Trig] key, the
sweep points are swept one by one.
Figure 5-10
Sweep Mode
5.4.2
List Sweep Parameters Setup
The sweep parameter used in list sweep measurement can be measurement frequency and
signal level. Use the sweep point field to specify the list sweep measurement parameter.
To select the list sweep measurement parameter
Step 1
Enter the [LIST TABLE] page.
Step 2
Use the cursor key to select FREQ[Hz] or LEVEL[V] field;
Step 3
Use the soft keys to select list sweep parameter.
Soft key
Function
FREQ[Hz]
Uses frequency as the list sweep parameter
LEVEL[V]
Uses voltage as the list sweep parameter
5.4.3
Configure the Sweep Points
AT381x's List sweep measurement feature supports up to 10 sweep points as well as
measurement limit values. Each of sweep point can be turned on or off.
To configure the sweep points
Step 1
Enter the [LIST TABLE] page.
Step 2
Use the cursor key to select any sweep point from 1~9;
Step 3
Turn sweep points on or off or enter sweep point data (frequency or level)
Soft key
Function
ON
Turn on current point
OFF
Turn off current point
LIST MEAS
Enter <LIST MEAS> Page
5.4.4
Limit parameters [LMT] selection
AT381x can compare the primary parameter or the secondary parameter limit while sweeping,
and give a PASS or FAIL conclusion.
Procedures for limit setting
Step 1
Enter the [LIST TABLE] page.
Step 2
Use the cursor key to select [CMP] of any sweep point from 1~10; Note that the current
sweep point needs to be turned on.