AT381x User's Guide
18
parallel equivalent circuit is used for LC oscillation circuits.
2.2.3
Range
Use 9-range test, auto, lock and nominal.
Nominal range (Applent new term definition): The instrument automatically selects the best
range based on nominal value.
2.2.4
Measurement Speed
The instrument is divided into four speeds: slow speed, medium speed and fast speed.
FAST speed: 40 times / sec
MED speed: 20 times / sec
SLOW speed: 3 times / sec
At the same time, 1-256 programmable average times can be performed to improve reading
stability.
2.2.5
Trigger Mode
Internal, external, manual, and remote trigger.
2.2.6
Basic Accuracy
AT3818 0.05%
AT3816A 0.05%
AT3816B 0.1%
AT3817A 0.05%
AT3810A 0.05%
AT3817D 0.1%
Measurement Display Range
Table 2-2
AT381x Measurement display range
Parameter
Measurement display range
L
0.00001nH
~
9999.99H
C
0.00001pF
~
999.999mF
R
、
X
、
Z
0.00001
Ω
~
99.9999M
Ω
B
、
G
0.01nS
~
999.999S
D
0.00001
~
9.99999
Q
0.00001
~
99999.9
θ
d
-179.999
°~
179.999°
θ
r
-3.14159
~
3.14159
%
-999.999%
~
999.999%
2.3
Signal Source
2.3.1
Test Frequency
AT3818
:
10Hz
~
300kHz continuous test frequency
AT3816A
:
50Hz
~
200kHz continuous test frequency