119
Modbus
(
RTU
)
Command
During the execution of correction, can determine whether correction is completed by
reading the register data.
Send: 01 03 5000 0001 950A
Response: 01 03 02 FFFF B9F4
Returns FFFF, indicating that the correction failed
Note:
When performing correction, try not to read the correction status frequently. Continuous
interrupts may cause the instrument to fail to perform correction.
Since the correction time is fixed, it is recommended that after the correction command is
issued, after the host waits for the correction time to elapse, it will get the correction result.
12.6.2
Full-frequency Short-circuit Correction [5008H]
Writing 0001 to Register 5008 will cause the AT381x to begin an open-circuit full-frequency
correction.
Since the correction process takes a few seconds, during this time, any write operation will be
ignored and only the read operation will be open. After the correction is completed, the write
command is normally opened.
Correction status can be obtained by reading the 5008 register during correction execution or
after correction:
0000
Correction success
FFFF
Correction fail
0001
In correction
Write
Please write a fixed value to the 5008 register:
:
00 01
Send
:
01 10 5008 0001 02 0001 36DD
Response
:
01 10 5008 0001 910B
Read
During the execution of correction, can determine whether correction is completed by
reading the register data.
Send
:
01 03 5008 0001 950A
Response
:
01 03 02 FFFF B9F4
Returns FFFF, indicating that the correction failed
Note:
When performing correction, try not to read the correction status frequently. Continuous
interrupts may cause the instrument to fail to perform correction.
Since the correction time is fixed, it is recommended that after the correction command is
issued, after the host waits for the correction time to elapse, it will get the correction result.