AT381x User's Guide
126
Test frequency
K
f
Typical frequency (direct correction)
0
Atypical frequency (interpolation correction)
0.0003
Table 13-4
Cable length factor
Test signal level
Cable length
0m
1m
2m
≤
1.5V
rms
0
2.5
×
10
-4
(1+0.05f
m
)
5
×
10
-4
(1+0.05f
m
)
>
1.5V
rms
0
2.5
×
10
-3
(1+0.016f
m
)
5
×
10
-3
(1+0.05f
m
)
In the table, fm is the test signal frequency [kHz].
13.3
Performace Test
Each test shall be carried out under the following working conditions.
Warm-u
p time: ≥ 20 minutes.
After warming up, open circuit and short circuit correction are performed.
The AT381x range works at “AUTO” to select the correct measurement range.
This test is only included in the test of main specifications. Other parameters not listed, users
can test under the specified conditions according to the specifications listed in this manual. If
the test result is found to beyond scope, please contact our maintenance department
immediately for repair.
13.3.1
Devices and Equipment Used for Performance Test
Table 13-5
Devices and equipment used for performance testing
No.
Equipment Name
Technical Requirements
1
Standard
capacitor
100pF
0.02
%
Loss D known
1000pF
10000pF
10nF
0.1
µ
F
1uF
2
AC standard
resistor
10
Ω
0.02
%
100
Ω
1k
Ω
10k
Ω
100k
Ω
3
Standard
inductor
100
µ
H
0.02%
1mH
10mH
100mH
4
Frequency meter
(
0
~
1000
)
MHz
5
Digital multimeter
0.5
%