AT381x User's Guide
132
15.
Specification
This chapter includes the following information:
Specifications
Dimension
Accuracy is defined as meeting all of the following conditions.
Temperature: 23℃±5℃
Humidity:
≤
65% R.H.
Zeroing: Open and Short Correction
Warm up time: >60 minutes
A 1-year calibration cycle
Test signal level: 10%
Test frequency accuracy: 0.01%
Parameter Test Basic Accuracy: 0.05%
15.1
General Specification
Display
:
True color TFT-LCD, Size: 3.5”
Test Function
:
Cs-Rs, Cs-D, Cp-Rp, Cp-D, Lp-
Rp, Lp-Q, Ls-Rs, Ls-Q,
Rs-Q, Rp-Q, R-X, DCR,
Z-
θr, Z
-
θ
d, Z-D, Z-Q
Monitor Parameter
:
(
2 sets
)Z, D, Q, Vac, Iac, Δ, Δ%, θr, θd,R, X, G, B, Y
Basic Accuracy
:
AT3818/AT3816A/AT3817A/AT3810A
0.05%
(
within basic range
)
Test Frequency
:
AT3818
:
10Hz
~
300kHz continuous test frequency
AT3816A
:
50Hz
~
200kHz continuous test frequency
AT3816B
:
50Hz
~
200kHz 37 points
AT3817A:
50Hz
~
100kHz continuous test frequency
AT3810A:
10Hz
~
20kHz continuous test frequency
AT3817D: 50Hz
~
100kHz, 10 points
Frequency range(F)
Resolution
10.0000Hz
≤
F
≤
99.9999Hz
0.0001Hz
100.0000Hz
≤
F
≤
999.999Hz
0.001Hz
1.00000kHz
≤
F
≤
9.99999kHz
0.01Hz
10.0000kHz
≤
F
≤
99.9999kHz
0.1Hz
100.000kHz
≤
F
≤
300.000kHz
1Hz
10.0000kHz
≤
F
≤
99.9999kHz
0.1Hz
100.000kHz
≤
F
≤
300.000kHz
1Hz
Frequency Accuracy
:
0.01%