R7
- OPERATING INSTRUCTION
7.2
S.H.S./S system
( Semplified Hygienization Sistem )
Models R7 CONTINENTAL, R7 INTERNATIONAL and R7 CART.
Description of the system
The system S.H.S. /S is equipped with a distilled water tank (
a
).
The tank can hold 1.8 liters.
Distilled water is delivered to:
• the sprays of all the instruments found on the instrument and assistant’s
board.
• the syringe on the assistant’s board.
• to fill the cup
• water quick-connect coupler (if present)
The icon button
on the TOUCH DISPLAY (see paragraph 5.1.1.2.8.)
allows activating/deactivating the SHS/S system.
NOTE:
the distilled water delivery status is shown by the icon (
A
)
on the control panel display.
Tank reserve level.
When the liquid in the tank falls below the reserve level, the relative icon
(
B
) appears on the instrument board’s console.
Filling the tank.
When the water level in the tank is low (about 500 cc), fill it as directed below:
• Deactivate the S.H.S. /S system by tapping the icon button
.
Check that the icon (
B
) on the console display disappears.
NOTE: during this operation, the pressurized air contained
in the tank will automatically be discharged to the outside.
• Turn the tank counter-clockwise (
a
) and remove it.
• Pour distilled water into the tank until the maximum level is reached.
WARNING!
Use only distilled water. For a higher guarantee of hygiene you can
add 600 parts per million (ppm) of hydrogen peroxide using 20 ml of
Peroxy Ag+ (not available for the American and Canadian markets) per
litre of distilled water, or oxygenated water (20 ml of 3% oxygenated
water per litre of distilled water).
• Put the tank back in place turning it clockwise.
WARNING!
Make sure that the tank is properly tightened.
• Touch the icon button
to re-enable the S.H.S. /S system and
confirm that filling is complete.
Check that the icon (
A
) appears on the console display.
WARNING!
If you are going to be absent from the surgery for long periods of time
(holidays), completely empty out the tank ( a ) before leaving.
Cleaning the tank.
We suggest cold-sterilizing periodically (at least once a month) only the tank
with peracetic acid-based products (for American and Canadian markets
we recommend using an EPA-approved product) in the following manner:
• take the tank out of the dental unit and empty it;
• prepare the solution of peracetic acid based product following the instruc-
tions provided by the manufacturer;
• fill the tank up to the rim with the peracetic acid based product;
• let the peracetic acid solution soak in the tank for the time stated by the
manufacturer;
• empty the peracetic acid solution from the tank;
• rinse the tank with distilled water;
• fill the tank with distilled water and, if necessary, add hydrogen peroxide
or oxygenated water as described above;
• put the tank back in place in the dental unit.
Содержание SMART TOUCH R7
Страница 1: ...R7 SMART TOUCH 97050649 rev 006 10 2016 EN...
Страница 2: ......
Страница 81: ...81 81 EN R7 OPERATING INSTRUCTION 11 1 Overall dimensions R7 CONTINENTAL...
Страница 82: ...82 82 EN R7 OPERATING INSTRUCTION 11 2 Overall dimensions R7 INTERNATIONAL...
Страница 83: ...83 83 EN R7 OPERATING INSTRUCTION 11 3 Overall dimensions R7 CART...
Страница 84: ...84 84 EN R7 OPERATING INSTRUCTION 11 4 Overall dimensions R7 M CONTINENTAL...
Страница 85: ...85 85 EN R7 OPERATING INSTRUCTION 11 5 Overall dimensions R7 M INTERNATIONAL...
Страница 86: ...86 86 EN R7 OPERATING INSTRUCTION 11 6 Overall dimensions R7 M CART...
Страница 87: ...87 87 EN R7 OPERATING INSTRUCTION 11 7 Overall dimensions R7 P...
Страница 88: ...88 88 EN R7 OPERATING INSTRUCTION 11 8 Overall dimensions R7 SINGLE HYDRO GROUP...
Страница 90: ......