46
46
EN
R7
- OPERATING INSTRUCTION
5.5.1.
RESTORATIVE operating mode
RESTORATIVE operation characteristics
- speed adjustable from 100 to 40000 RPM (handpiece 1:1 ),
- Rotation speed change mode settable from variable to fixed and vice versa
- Alarm signal when the maximum torque is reached
- Fast capture of the maximum speed during motor rotation.
Menu with micromotor extracted but not active.
All the icon buttons are active and each function available can be changed
(see paragraph 5.5.).
NOTE:
each setting or value changed will automatically be stored
in the operating program selected (e.g. P1).
Menu with micromotor extracted and active.
The modifiable functions are the following:
• Maximum drill rotation speed using the icon buttons
or
,
• Current speed freezing using the following icon button:
Sets the current rotation speed as maximum speed
• Foot control lever change mode using the following icon buttons:ù
Sets the current rotation speed as maximum speed
at the same time activating a function to change
the foot control lever ON/OFF mode
Switches the foot control lever change mode from
ON/OFF to linear
5.5.2.
ENDODONTIC operating mode
ENDODONTIC operation characteristics
- Speed adjustable from 100 to 600 rpm with the value always referring to the
drill irrespective of the reduction ratio
- Torque adjustable from 0.1 to 5.0 Ncm, excluding the 1:1 reduction gear (4.5
Ncm)
- Motor rotation speed change mode settable from variable to fixed and vice
versa
- Fast capture of the maximum speed during motor rotation.
Menu with micromotor extracted but not active.
All the icon buttons are active and each function available can be changed
(see paragraph 5.5.).
As well as the standard settings, in ENDODONTIC mode you can also set
“Operation when maximum torque reached” by touching the relative icon
button:
Rotation lock
Rotation lock and subsequent inversion of the rotation direction
Rotation lock, inversion of the normal rotation direction and subse-
quent return to the normal rotation direction
NOTE:
each setting or value changed will automatically be stored
in the operating program selected (e.g. P1).
Содержание SMART TOUCH R7
Страница 1: ...R7 SMART TOUCH 97050649 rev 006 10 2016 EN...
Страница 2: ......
Страница 81: ...81 81 EN R7 OPERATING INSTRUCTION 11 1 Overall dimensions R7 CONTINENTAL...
Страница 82: ...82 82 EN R7 OPERATING INSTRUCTION 11 2 Overall dimensions R7 INTERNATIONAL...
Страница 83: ...83 83 EN R7 OPERATING INSTRUCTION 11 3 Overall dimensions R7 CART...
Страница 84: ...84 84 EN R7 OPERATING INSTRUCTION 11 4 Overall dimensions R7 M CONTINENTAL...
Страница 85: ...85 85 EN R7 OPERATING INSTRUCTION 11 5 Overall dimensions R7 M INTERNATIONAL...
Страница 86: ...86 86 EN R7 OPERATING INSTRUCTION 11 6 Overall dimensions R7 M CART...
Страница 87: ...87 87 EN R7 OPERATING INSTRUCTION 11 7 Overall dimensions R7 P...
Страница 88: ...88 88 EN R7 OPERATING INSTRUCTION 11 8 Overall dimensions R7 SINGLE HYDRO GROUP...
Страница 90: ......