- OPERATING INSTRUCTION
4.
Dental chair operation
The dental chair can be moved as follows:
• Chair seat up/down,
• Back up/down with inclination of the chair seat (Tren-
delemburg compensated),
• 30° clockwise or counter-clockwise seat rotation
(optional).
The dental chair can be operated from the following
places:
• Instrument board (
a
) (see par. 5).
• Multifunction foot control (
b
) (see par. 5.2).
• Assistant’s board (
c
) (see par. 6).
• Dental chair side control panel (see par. 4.4)
Dental chair movement shutdown
With the instruments in rest position, you can di-
sable the dental chair movements (see paragraph
5.1.1.2.5.).
The movement disabling is shown on the control
panel display by the relevant icon (
A
).
3.2.
R7 M line operating unit conversion
To convert an R7 M line operating unit from the right-handed operator
version to the left-handed operator version, simply invert instrument’s table
arm position with operating lamp support arm after having removed the
anti-rotation stopper pin (
c
). Once both arms have been turned, insert anti-
rotation stopper pin (
c
) back in place to lock the operating lamp support
arm in the new position.
4.1.
Safety devices
The equipment is supplied with the following safety devices:
• Dental chair footboard is equipped with a device (
l
) that immediately
stops the dental chair down movement in the presence of an obstacle
and automatically moves the seat up to free the obstacle.
• The dental chair back rest is equipped with a device (
m
) which immediately
stops the back rest or dental chair from moving down when an obstacle is
encountered and automatically moves it back up to clear the obstacle.
• The support arms of the assistant’s module feature a safety device (
n
)
which stops downward chair movement if an obstacle is encountered. The
chair will then automatically move upwards so the user can remove the
obstruction.
• The hydrogroup support arms are equipped with a safety device (
o
)
which immediately stops the dental chair from moving down when an
obstacle is encountered and automatically moves it back up to clear the
obstacle.
• Dental chair movements:
- with the instrument extracted NOT working: manual movements allowed,
automatic movements inhibited, but if they are already in progress at
the moment of extraction they are not interrupted;
- with instrument extracted and working: all the dental chair movements
are inhibited.
Содержание SMART TOUCH R7
Страница 1: ...R7 SMART TOUCH 97050649 rev 006 10 2016 EN...
Страница 2: ......
Страница 81: ...81 81 EN R7 OPERATING INSTRUCTION 11 1 Overall dimensions R7 CONTINENTAL...
Страница 82: ...82 82 EN R7 OPERATING INSTRUCTION 11 2 Overall dimensions R7 INTERNATIONAL...
Страница 83: ...83 83 EN R7 OPERATING INSTRUCTION 11 3 Overall dimensions R7 CART...
Страница 84: ...84 84 EN R7 OPERATING INSTRUCTION 11 4 Overall dimensions R7 M CONTINENTAL...
Страница 85: ...85 85 EN R7 OPERATING INSTRUCTION 11 5 Overall dimensions R7 M INTERNATIONAL...
Страница 86: ...86 86 EN R7 OPERATING INSTRUCTION 11 6 Overall dimensions R7 M CART...
Страница 87: ...87 87 EN R7 OPERATING INSTRUCTION 11 7 Overall dimensions R7 P...
Страница 88: ...88 88 EN R7 OPERATING INSTRUCTION 11 8 Overall dimensions R7 SINGLE HYDRO GROUP...
Страница 90: ......